電話:0755-82515004
手機(jī):13760205028
公司地址:深圳市福田區(qū)深南中路3037號(hào)南光捷佳大廈Email:liu56817@126.com
QQ:528642442
Vermason 222547校準(zhǔn)器 Calibration Unit for Vermason Wrist Strap and Footwear
Tester
用于校準(zhǔn)手腕帶和工鞋接地測(cè)試儀
適配儀器:222504, 222506,222507, 222562, 222563, 222564, 222565
Vermason(DESCO子品牌)
Operation
The Tester Calibration Unit was specifically designedfor use in calibration of
Vermason brand equipment.The following procedures will not cover adjustment of
the test equipment. For detailed information regardingadjustment of specific
Vermason brand testers, pleasecontact Desco Europe Customer Service.
Frequency of recalibration should be based on thecritical nature of those ESD
sensitive items handled andthe risk of failure for the ESD protective equipment
andmaterials. In general, we recommend that calibration be
performed annually.
Testing the Compact Tester
WRIST STRAP CIRCUIT
1. Connect the 10 mm lead from the Calibration Unitto the wrist strap stud on
the Compact Tester.
2. Select 715K FAIL LOW on the Calibration Unit.
3. Press and hold the Calibration Unit’s test probeon the Compact Tester’s test
plate. The testershould indicate a FAIL LOW condition.
4. Select 785K PASS LOW on the Calibration Unitand repeat the test. The tester
should indicate aPASS condition.
5. Select 33M PASS HIGH on the Calibration Unitand repeat the test. The tester
should indicate aPASS condition.
6. Select 37M FAIL HIGH on the Calibration Unitand repeat the test. The tester
should indicate aFAIL HIGH condition.
FOOTWEAR CIRCUIT
1. Connect the 10 mm lead from the Calibration Unitto the 10 mm footwear stud on
the Compact Tester.
2. Select 715K FAIL LOW on the Calibration Unit.
3. Press and hold the Calibration Unit’s test probeon the Compact Tester’s test
plate. The testershould indicate a FAIL LOW condition.
4. Select 785K PASS LOW on the Calibration Unitand repeat the test. The tester
should indicate aPASS condition.
5. Select 33M PASS HIGH on the Calibration Unitand repeat the test. The tester
should indicate aPASS condition.
6. Select 37M FAIL HIGH on the Calibration Unit andrepeat the test. The tester
should indicate aFAIL HIGH condition